Short tau inversion recovery
Short tau inversion recovery (STIR), also known as short TI inversion recovery, is a fat suppression technique with an inversion time TI = ln(2)·T1fat,
where the signal of fat is zero. This equates to approximately 140 ms at 1.5 T.
To distinguish two tissue components with this technique, their T1 values must be different. FLAIR is a similar technique to suppress water.
Inversion recovery imaging allows homogeneous and global fat suppression and can be used with low-field-strength magnets. However, this technique is not specific for fat. Thus, the signal intensity of tissue with a long T1 and tissue with a short T1 may cause ambiguity.
Since STIR sequences use short inversion recovery time, they cannot be used with gadolinium injection because tissues that take up gadolinium will exhibit T1 shortening and may inadvertently be nulled.
Related Radiopaedia articles
Physics and Imaging Technology: MRI
- MRI (introduction)
-
MRI physics
- B0
- chemical shift
- dependence of magnetization (proton density, field strength and temperature)
- echo time
- eddy currents
- electromagnetic induction
- Ernst angle
- flip angle
- Larmor frequency
- magnetic dipole magnetic field gradient
- magnetic susceptibility
- magnetism
- molecular tumbling rate effects on T1 and T2
- net magnetization vector (NMV)
- relaxation
- repetition time
- resonance and radiofrequency (RF)
- units of magnetism
- MRI hardware
- signal processing
-
MRI pulse sequences (basics | abbreviations | parameters)
- CSF flow studies
- diffusion weighted sequences (DWI)
- echo-planar pulse sequences
- fat-suppressed imaging sequences
- gradient echo sequences
- inversion recovery sequences
- metal artifact reduction sequence (MARS)
-
perfusion-weighted imaging
- techniques
- derived values
- saturation recovery sequences
- spin echo sequences
- spiral pulse sequences
- susceptibility-weighted imaging (SWI)
- T1 rho
- MR angiography (and venography)
-
MR spectroscopy (MRS)
- 2-hydroxyglutarate peak: resonates at 2.25 ppm
- alanine peak: resonates at 1.48 ppm
- choline peak: resonates at 3.2 ppm
- citrate peak: resonates at 2.6 ppm
- creatine peak: resonates at 3.0 ppm
- functional MRI (fMRI)
- gamma-aminobutyric acid (GABA) peak: resonates at 2.2-2.4 ppm
- glutamine-glutamate peak: resonates at 2.2-2.4 ppm
- Hunter's angle
- lactate peak: resonates at 1.3 ppm
- lipids peak: resonates at 1.3 ppm
- myoinositol peak: resonates at 3.5 ppm
- MR fingerprinting
- N-acetylaspartate (NAA) peak: resonates at 2.0 ppm
-
MRI artifacts
- MRI hardware and room shielding
- MRI software
- patient and physiologic motion
- tissue heterogeneity and foreign bodies
- Fourier transform and Nyquist sampling theorem
- MRI contrast agents
- MRI safety