Slice-overlap artifact
The slice-overlap artifact, also known as cross-talk artifact, is a name given to the loss of signal seen in an image from a multi-angle, multi-slice acquisition, as is obtained commonly in the lumbar spine. It should not be confused with cross excitation which although similar in causation, is not due to angled images.
If the slices obtained at different disk spaces are not parallel, then the slices may overlap. If two levels are done at the same time, e.g. L4-5 and L5-S1, then the level acquired second will include spins that have already been saturated. This causes a band of signal loss crossing horizontally in your image, usually worst posteriorly. The dark horizontal bands in the bottom of the following axial image through the lumbar spine demonstrates this artifact.
As long as the saturated area stays posterior to the spinal canal it causes no harm.
Related Radiopaedia articles
Physics and Imaging Technology: MRI
- MRI (introduction)
-
MRI physics
- B0
- chemical shift
- dependence of magnetization (proton density, field strength and temperature)
- echo time
- eddy currents
- electromagnetic induction
- Ernst angle
- flip angle
- Larmor frequency
- magnetic dipole magnetic field gradient
- magnetic susceptibility
- magnetism
- molecular tumbling rate effects on T1 and T2
- net magnetization vector (NMV)
- relaxation
- repetition time
- resonance and radiofrequency (RF)
- units of magnetism
- MRI hardware
- signal processing
-
MRI pulse sequences (basics | abbreviations | parameters)
- CSF flow studies
- diffusion weighted sequences (DWI)
- echo-planar pulse sequences
- fat-suppressed imaging sequences
- gradient echo sequences
- inversion recovery sequences
- metal artifact reduction sequence (MARS)
-
perfusion-weighted imaging
- techniques
- derived values
- saturation recovery sequences
- spin echo sequences
- spiral pulse sequences
- susceptibility-weighted imaging (SWI)
- T1 rho
- MR angiography (and venography)
-
MR spectroscopy (MRS)
- 2-hydroxyglutarate peak: resonates at 2.25 ppm
- alanine peak: resonates at 1.48 ppm
- choline peak: resonates at 3.2 ppm
- citrate peak: resonates at 2.6 ppm
- creatine peak: resonates at 3.0 ppm
- functional MRI (fMRI)
- gamma-aminobutyric acid (GABA) peak: resonates at 2.2-2.4 ppm
- glutamine-glutamate peak: resonates at 2.2-2.4 ppm
- Hunter's angle
- lactate peak: resonates at 1.3 ppm
- lipids peak: resonates at 1.3 ppm
- myoinositol peak: resonates at 3.5 ppm
- MR fingerprinting
- N-acetylaspartate (NAA) peak: resonates at 2.0 ppm
-
MRI artifacts
- MRI hardware and room shielding
- MRI software
- patient and physiologic motion
- tissue heterogeneity and foreign bodies
- Fourier transform and Nyquist sampling theorem
- MRI contrast agents
- MRI safety